The use of vascular closure devices is widespread both with operators who carry out their interventions via the femoral approach and those using radial access. Interventionists performing arterial access procedures should be not only aware of when and how to use them but also of the advantages and disadvantages of each device, eventually predicting and treating any complications which may occur. In this chapter we aim to provide an overview of the currently available devices, focusing on how they work, on their efficacy in achieving haemostasis, on any risks associated with their use, and on the data available for each device. We also compare the published registries and meta-analyses and look at the current ACC/AHA guidelines for vascular closure. The majority of devices in this chapter are described as used for coronary and peripheral interventions. Vascular closure related to trancatheter aortic valve implantation is covered in detail in the chapter Transcatheter aortic valve implantation (old).
Percutaneous treatment for coronary artery disease has constantly expanded over the last two decades. With improvements in techniques and equipment, a large number of patients with ischaemic heart disease are undergoing invasive vascular interventions.
Approximately 7 million invasive vascular procedures are performed worldwide each year and this number is expected to increase further with an aging population. The majority of these procedures are performed via transradial approach but the femoral artery remains a commonly used route.
Interventional procedures may necessitate the use of larger sheath sizes than a diagnostic procedure. In addition, the use of heparin and potent antiplatelets treatments during percutaneous coronary intervention (PCI) makes achieving immediate haemostasis more challenging. For more complex interventions such as TAVI or the insertion of percutaneous left assist devices where larger arteriotomies are required, the femoral artery is the standard choice of access.
Vascular complications, which have been reported as high as 6% in some series, remain the leading cause of morbidity after cardiac catheterisation . A steadily increasing number of these procedures like complex coronary angioplasty and percutaneous aortic valve implantation are performed in high-risk patients through the femoral artery. Thus, it is essential to have effective arterial haemostasis devices in order to reduce vascular access complications.
Manual compression has traditionally been considered the “gold standard” approach for closure of the arteriotomy site requiring prolonged immobilisation and prolonged surveillance to achieve success. However, the duration of compression and the time to ambulation are generally proportional to the size of the introducer sheath and the amount of anticoagulants used. Manual compression is very effective for small sheath sizes but with larger sizes, the immobilisation time and the tendency towards complications appear to increase.
Since the development of vascular closure devices (VCDs) in the mid-1990s in order to improve patient comfort via a reduction in vascular complications, time to haemostasis and time to ambulation, this field has seen a rapid growth with the development of a multitude of devices incorporating sutures, collagen plugs, nitinol clips and other approaches to achieve haemostasis. In an era of cost effective healthcare, there has been a growing interest in technologies that can speed vascular closure allowing earlier ambulation and shorter hospital stay. It has been recently reported that the global market for vascular closure devices was estimated at $ 0.9 billion in 2021. It is predicted that global sales figures will have reached an estimated $1.3 billion by 2026.
Despite the widespread use of VCDs, there is a paucity of data on their safety and efficacy. In this chapter, we will review the devices currently available on the market, the rationale and the evidence base for VCDs.
See chapter Vascular access
Diagnostic and interventional cardiovascular procedures are routinely performed via the common femoral or the radial artery. Usage of the radial approach is increasing worldwide and in many European countries it is considered routine practice. The ACUITY trial showed a reduction in major bleeding complications in patients with acute coronary syndromes managed invasively when the transradial approach was used and this has been consolidated by many meta-analyses , . This is most likely due to the relative ease of compression of the radial artery.
However, the transfemoral route remains widely used, especially in the United States where the transradial approach accounted for only 25.2% of the total procedures performed in 2014 . More contemporary data albeit in the context of a trial would suggest a prevalence of up to 60% . In addition, crossover from a trans-radial access to transfemoral is required in about 7 % of patients for various reasons.
Many patient- and procedure-related factors can increase the risk of major femoral bleeding while a non-rigorous puncture outside the safe location almost quadruples the risk of vascular complication. This is the reason why it is vitally important that appropriate techniques regarding identification, puncture and location of sheath insertion into the common femoral artery be used to reduce the occurrence of access site complications , . (Optimal location for femoral artery access. The outlined dot marks the sheath entry site into the common femoral artery (CFA) at the medial aspect of the femoral head in the midline. (A) Inferior border of the femoral head. (C) Line demarcating inguinal ligament running from the anterior superior iliac spine to the pubic symphysis. (D) Bifurcation of CFA into superficial femoral artery and profunda.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-1A.pptx"})' class="cursor-pointer">Figure 1A, Recommended approach: Using fluoroscopy as an adjunct in gaining femoral artery access. Left panel: diagramatic representation of angiographic landmarks. Right panel: puncture needle (arrow) can be seen entering the common femoral artery (CFA) at the optimal location in the medial portion of the midline of the femoral head.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-1B.pptx"})' class="cursor-pointer">Figure 1B, Femoral angiography: For a right femoral arteriotomy, the right anterior oblique (RAO) 30\/40 degree view (left) nicely lays out the femoral bifurcation and the level of the inferior epigastric artery (arrow), but obscures the sheath entry site. The left anterior oblique (LAO) 30\/40 degree view (right) displays the sheath entry site, but the bifurcation is obscured.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-1C.pptx"})' class="cursor-pointer">Figure 1C and Focus box 1).
  
  Optimal location for femoral artery access. The outlined dot marks the sheath entry site into the common femoral artery (CFA) at the medial aspect of the femoral head in the midline. (A) Inferior border of the femoral head. (C) Line demarcating inguinal ligament running from the anterior superior iliac spine to the pubic symphysis. (D) Bifurcation of CFA into superficial femoral artery and profunda.
  
  Recommended approach: Using fluoroscopy as an adjunct in gaining femoral artery access. Left panel: diagramatic representation of angiographic landmarks. Right panel: puncture needle (arrow) can be seen entering the common femoral artery (CFA) at the optimal location in the medial portion of the midline of the femoral head.
  
  Femoral angiography: For a right femoral arteriotomy, the right anterior oblique (RAO) 30/40 degree view (left) nicely lays out the femoral bifurcation and the level of the inferior epigastric artery (arrow), but obscures the sheath entry site. The left anterior oblique (LAO) 30/40 degree view (right) displays the sheath entry site, but the bifurcation is obscured.
Ultrasound (US) guided cannulation of the common femoral artery may reduce vascular complications by delimiting the optimal zone where to puncture as well as determining the localization of parietal calcifications of the common femoral artery. In many randomized trials, US guidance has been shown to reduce complications, number of attempts, and the time needed for vascular access . Its routine use has not yet been recommended in current guidelines for interventional vascular procedures. However, in the light of available data, it should be considered for any patient at high risk of complications, or with difficult access, most notably for large-bore access.
Moreover, the femoral angiography at the end of the procedure allows clarification of the (approximately) 13% of patients with unsuitable anatomy for the usage of VCDs and also allows identification of sheath insertion above the inferior epigastric artery which is a major risk factor for retroperitoneal haemorrhage , . Identification of significant femoral arterial disease allows selective use of manual compression in situations where insertion of a VCD may be associated with an increased rate of complications. Obviously, the combination of fluoroscopy with US guidance may decrease the likelihood of a high puncture if the US probe inadvertently slides cranially.
See chapter The prevention and management of complications during percutaneous coronary intervention
Femoral complications following vascular access vary from 2-6% following interventional procedures (Femoral access complication rates <\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_TABLE-1.pptx"})' class="cursor-pointer">Table 1) . Vascular complications increase with increasing complexity of the procedure and there are many identifiable predictors during PCI (see Focus box 2). A multicentre registry in northern New England including over 18,000 patients demonstrated a vascular complication rate of 3% following PCI. A multivariate analysis of this data showed recognisable risk factors to be age >70 years, females, body surface area <1.6 m2, renal failure, urgent procedures, complex disease and use of glycoprotein IIb/IIIa inhibitors . These patients displayed features, which enabled them to be categorised as low risk, moderate risk or high risk in terms of potential for complications (Clinical determinants of vascular complication rates<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_TABLE-2.pptx"})' class="cursor-pointer">Table 2).
  
  Femoral access complication rates
  
  Clinical determinants of vascular complication rates
To try to reduce the incidence of these complications, VCDs have been introduced as adjuncts or alternatives to manual compression for haemostasis. Since their development, rates of major complications following PCI have been noted to be closer to 2% , , representing a 42% relative decrease in complications from 2002-2006 . Given that the characteristics of patients undergoing PCI have not changed over the last decade, it can be hypothesised that the reason for the decrease may reflect better patient selection for VCDs, improved femoral access techniques, changes in PCI pharmacology, improved VCDs and better operator experience with VCD use , .
Vascular closure devices were first developed in the mid-1990s. They were invented out of a necessity to achieve rapid and successful haemostasis in patients undergoing vascular interventions in the era of increased anticoagulation and increasingly complex procedures.
The first device (VasoSeal; Datascope Corporation, Montvale, NJ, USA) was a collagen-based device and the only VCD significantly associated with an increased risk of vascular complications when compared with manual compression in 2 meta-analyses , . Subsequent technologies used active closure devices including sutures alone , extravascular collagen and combinations of collagen and suture material . The learning curve for each device was steep and complication rates were higher in the early period of VCD usage.
Since this time, the original devices have undergone significant improvements while maintaining their initial concepts. In addition, new devices incorporating a staple/clip technique, a synthetic biosealant-based system, passive closure devices focusing on enhanced manual compression using external patches with prothrombotic coatings, wire-stimulated track thrombosis or assisted compression with mechanical clamps, have been developed. Their usage have resulted in a shorter time to haemostasis, earlier patient ambulation, high procedural success rates and improved patient satisfaction , , .
VCDs can be classified as either active or passive devices according to the method of closure. In active devices, the arteriotomy site is closed by mechanically bringing the margins closer via either suture devices, clips, or collagen plug devices. In the case of passive closure devices, a plug, sealant, or gel is deployed at the arteriotomy site without physically closing the arteriotomy. The classification is shown in Classification of VCDs<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_TABLE-3.pptx"})' class="cursor-pointer">Table 3.
  
  Classification of VCDs
The Angio-Seal™ family (Terumo Medical Corporation, Somerset, NJ, USA) consists of three devices - Angio-Seal™ VIP, Angio-Seal™ STS Plus and Angio-Seal™ Evolution. The core principle of the devices involves sealing the arteriotomy site using an anchor on the inside of the artery and a collagen plug on the outside both of which are linked by a suture. Haemostasis is achieved primarily through this mechanical seal and is supplemented by the coagulation-inducing properties of the collagen. All three structures are completely bio-absorbed by 3 months of implantation. The devices are available in 6Fr and 8Fr configurations. The 6Fr Angio-Seal device is suitable for closure after usage of a 6 Fr or smaller procedural sheath (minimal artery diameter of 4mm), while the 8Fr Angio-Seal device is for closure after usage of an 8 Fr or smaller procedural sheath. All three structures are completely bio-absorbed by 3 months of implantation. Once an Angio-Seal™ is deployed, re-puncture in the same artery is possible within 90 days but must occur either proximal or distal to the deployed device.
The Angio-Seal™ STS Plus Platform features a self-tightening suture, which eliminates the need for the post placement spring used in older models (Angio-Seal\u2122 STS device; Assembled 6Fr (green) and 8Fr (blue) insertion sheath and carrier tube<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-2.pptx"})' class="cursor-pointer">Figure 2). This feature allows the arterial closure procedure to be completed in the catheterisation room. . The device creates a mechanical seal by sandwiching the arteriotomy between a bioabsorbable anchor and a collagen sponge, which dissolves within 60-90 days (Angio-Seal\u2122 STS device(A) Undeployed Angio-Seal\u2122 demonstrating the absorbable collagen sponge and the absorbable polymer anchor, connected by an absorbable self-tightening suture (STS).(B) Deployed Angio-Seal\u2122 demonstrating the compressed collagen sponge and anchor forming a tight compressive sandwich.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-3.pptx"})' class="cursor-pointer">Figure 3). Once an Angio-Seal™ is deployed, re-puncture in the same artery is possible within 90 days but must occur either proximal or distal to the deployed device.
  
  Angio-Seal™ STS device; Assembled 6Fr (green) and 8Fr (blue) insertion sheath and carrier tube
  
  Angio-Seal™ STS device(A) Undeployed Angio-Seal™ demonstrating the absorbable collagen sponge and the absorbable polymer anchor, connected by an absorbable self-tightening suture (STS).(B) Deployed Angio-Seal™ demonstrating the compressed collagen sponge and anchor forming a tight compressive sandwich.
The Angio-Seal™ procedure is composed of three stages:
(A) locating the artery;
(B) setting the anchor;
(C) sealing the puncture.
  
  The Angio-Seal™ insertion sheath and vessel locator assembly inserted into the femoral artery over a guidewire
  
  The Angio-Seal™ device (insertion sheath and carrier tube) is withdrawn until the anchor is seated over the artery puncture
  
  The Angio-Seal™ collagen plug is pushed gently into the tissue tract using the tamper tube
  
  The Angio-Seal™ collagen plug twists as it is pushed into the tissue tract
The Angio-Seal™ VIP device was released onto the international market in 2006. Due to a V-Twist Integrated Platform technology, this device provides a larger collagen footprint. The device features a Bondek® Plus Suture (Teleflex Medical, Limerick, PA, USA) with a patented Polyglyd™ coating (Teleflex Medical) for enhanced lubricity. The lubricious suture enables smooth compaction for a uniform and secure seal, while maintaining secure knot locking ability. It is deployed using the same technique as for the STS system.
Consistent with its Angio-Seal™ predecessors, Angio-Seal™ Evolution achieves haemostasis through the deployment of an anchor, suture and collagen seal (Angio-Seal\u2122 Evolution<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-5.pptx"})' class="cursor-pointer">Figure 5). With Evolution, however, single-handed deployment is possible, which enables physicians to support the puncture site more easily. The device's automated collagen compaction system ensures consistent compaction of the collagen against the exterior wall of the vessel reducing procedural variables.
  
  Angio-Seal™ Evolution
The Angio-Seal™ pivotal randomised, controlled, multicentre trial was published in 1995 and proved the device was safe and efficacious in a group of 435 patients randomised in a 1:1 fashion to the Angio-Seal device or manual compression . Time to haemostasis was considerably shorter (2.5 versus 15.3 minutes, p<0.0001) with a lower complication rate. Subgroup analysis showed benefit in those patients undergoing interventional procedures.
Use of the Angio-Seal has been associated with high success rates in the range of 91-100% , , , , . The 6 Fr device was evaluated in a prospective multicentre registry and was proven to be safe providing immediate haemostasis in 91% of patients .
The CAP trial compared the Angio-Seal VIP with the Perclose ProGlide and to manual compression with respect to time to haemostasis and ambulation, patient satisfaction, and vascular complications following PCI in 200 patients . The Angio-Seal was deployed in 100% of patients. The Angio-Seal VIP allowed for earlier haemostasis and ambulation compared with both compression and Perclose ProGlide and was associated with greater patient satisfaction compared with compression (p<0.01).
The Angio-Seal Evolution was evaluated in a multicentre registry involving 10 sites and over 1,000 patients and demonstrated safety and efficacy after routine cardiac catheterisation and PCI .
Although the effects of Angio-Seal have proved promising, some studies have shown preference for the FemoStop™device . The complication and failure rates of Angioseal and the Mynx® were compared in an all-comers population undergoing percutaneous coronary intervention between 2008 and 2014 . 4,074 patients were included, 2,910 receiving the Angioseal and 1,164 receiving Mynx. VCD choice was at the operator’s discretion. Safety was assessed by vascular complications defined as either vascular injury or access-site bleed (haemoglobin drop >3g/dL requiring transfusion, retroperitoneal bleed, or haematoma >5cm, or the composite of both). Efficacy was evaluated by device failure and defined as inability to achieve immediate haemostasis or use of additional haemostatic mechanisms. Outcomes were evaluated at 30-days. The groups were comparable with exceptions as follows: the Angio-Seal group was slightly younger (64±12 vs 65±12, p<0.001), with less peripheral arterial disease (11.3% vs 13.9%, p=0.03) and increased 7Fr sheath use compared with Mynx (59% vs 22%, p<0.001). Safety and efficacy were similar between both groups.
The Perclose® system (Abbott Vascular, Redwood City, CA, USA), introduced in 1994, was the first suture-mediated device to be approved by the Food and Drug Administration. It offers improvements in the ease of knot delivery and strength and in the non-inflammatory nature of the suture material. It is designed to deliver polyester suture to close femoral artery puncture sites following diagnostic or interventional procedures. It has received approval for closing arteriotomy sites ranging from 5Fr to 21Fr. The use of larger sheaths >8Fr requires a pre-close with two devices.
The Perclose® SMC device is composed of a sheath and a guide. The guide houses the needles and the foot precisely controls the placement of the needles around the puncture site. A marker lumen is contained within the guide, with the intraluminal port of the lumen positioned at the distal end of the needle guide. Proximally, the marker lumen exits from the handle of the device. The marker lumen allows a pathway for back bleeding from the artery to ensure proper device positioning. The device is compatible with a standard 0.038” (or smaller) guidewire.
A knot pusher accessory is included, and is designed to position the tied suture knot to the arteriotomy. The Perclose® Suture Trimmer is also designed to trim the trailing limbs of suture.
The Perclose® system is depicted in Figure 6 and Perclose\u00ae vascular closure<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-7.pptx"})' class="cursor-pointer">Figure 7.
  
  
  
  Perclose® vascular closure
Released in 2021, Prostyle is the latest generation design of the Perclose system. It is mainly characterized by higher tensile-strength needles, which both minimize needle deflection and improve penetration through complex and less compliant anatomy. In addition, a hydrophilic coating has been added to the Perclose device as well as to the sheath of suture trimmer, thus facilitating both device and knot advancement. Though it shares the same indications as Perclose Proglide and is deployed using the same technique, it is nonetheless marked by a higher rate of reliability and performance given its improved characteristics.
See chapter Transcatheter aortic valve implantation (old)
The Prostar® XL Percutaneous Vascular Surgical (Abbott Vascular, Redwood City, CA, USA) device is designed to deliver polyester suture(s) to close femoral artery puncture sites following catheterisation procedures. The Prostar XL is indicated for closure of femoral arterial access in patients who have undergone procedures using 8.5 Fr to 10 Fr sheaths. It can also be used for pre-closure of large arteriotomies.
The Prostar XL 10 Fr device consists of a J-shaped hydrophilic sheath (Label A - The Abbott Prostar\u00ae XL Percutaneous Suture-Mediated Closure Device. (A) Prostar XL 10F device. (B) Knot pusher and inner components showing the needles with respect to the suture and needle guide.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-8.pptx"})' class="cursor-pointer">Figure 8) which contains two pairs of sutured needles, a needle guide which precisely controls the needles around the puncture site, and a rotating barrel which receives deployed needles and a marker lumen (Label B - The Abbott Prostar\u00ae XL Percutaneous Suture-Mediated Closure Device. (A) Prostar XL 10F device. (B) Knot pusher and inner components showing the needles with respect to the suture and needle guide.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-8.pptx"})' class="cursor-pointer">Figure 8). Proximally, the marker lumen exits from the hub of the device and provides a pathway for back bleeding from the femoral artery and ensures proper device positioning. The barrel rotates independently from the central core and is designed to prepare the subcutaneous track via blunt dissection. Barrel rotation is accomplished by depressing the interlocks exiting from the hub. The Prostar XL devices track over a standard 0.038” (or smaller) guidewire.
  
  The Abbott Prostar® XL Percutaneous Suture-Mediated Closure Device. (A) Prostar XL 10F device. (B) Knot pusher and inner components showing the needles with respect to the suture and needle guide.
Other additional accessories are the Perclose® Knot Pusher, which is used to advance the tied knot down to the arteriotomy.
The Perclose Arterial Tamper can be used to augment haemostasis by positioning the knot over the surface of the artery (Labels A-C - Deployment of the Abbott Prostar\u00ae XL Percutaneous Suture-Mediated Closure Device(A) Schematic of the Prostar XL device, maintaining a 45-degree angle during deployment; (B) Cartoon depiction of device deployment showing the green and white sutures penetrating the arterial wall; (C) Drawing of the green and white sutures as they penetrate the arterial wall following deployment.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-9.pptx"})' class="cursor-pointer">Figure 9).
  
  Deployment of the Abbott Prostar® XL Percutaneous Suture-Mediated Closure Device(A) Schematic of the Prostar XL device, maintaining a 45-degree angle during deployment; (B) Cartoon depiction of device deployment showing the green and white sutures penetrating the arterial wall; (C) Drawing of the green and white sutures as they penetrate the arterial wall following deployment.
Suture management
The “Closer” devices can be used with several different sheath sizes. “Perclose™” can be performed prior to sheath insertion when using very large sheaths, allowing the operator to create the ‘’purse-string’’ before actually dilating the arteriotomy. This is especially useful in percutaneous aortic valve implantations and closures of abdominal aortic aneurysms, where sheath sizes currently approach 18 Fr.
The major trials for the Perclose® device were the STAND I and STAND II trials. The STAND I trial (a non-randomised registry) evaluated the device in 200 patients undergoing diagnostic procedures with successful haemostasis in 99% of patients in a median time of 13 minutes and with a major complication rate of 1% .
The STAND II trial was the pivotal randomised controlled multicentre trial . 515 patients undergoing diagnostic or interventional procedures were randomised to the use of the 8 Fr or the 10 Fr Prostar-Plus device versus manual compression. Successful haemostasis was achieved in 97.6% of patients compared with 98.9% by compression (p=ns). In the diagnostic cases, the time to discharge was significantly reduced from 8.3 hours in the compression arm to 4.4 hours in the Prostar arm. In the interventional cases, time to discharge remained unaltered. There was a slight trend towards increased complications with the suture-based device but this was not statistically significant. The STAND II trial proved the safety and efficacy of a suture-based concept for arteriotomy closure.
The safety and efficacy of the Perclose® has been supported by the results of other randomised studies , , , , , , .
The newest generation device was evaluated in 72 procedures with a success rate of 98.6% without any notable complications . Immediate ambulation with this device has been shown to be feasible and safe .
A useful feature of the Perclose® system is the ability for "pre-closure" . The sutures are deployed at the beginning of the procedure using a smaller sheath, which is then exchanged for the Perclose® device. A larger sheath is then placed and the sutures tied after completion of the procedure. This method has been used successfully with 7-8 Fr sheaths as well as with larger , , , , , , Fr sheaths , , used for percutaneous transaortic valve implantations and endoluminal repair of abdominal aortic aneurysms.
Reports have shown safety when closing up to 22 Fr femoral arteriotomies and brachial access sites . The device can be used several times within days and the device can be used multiple times within the same femoral artery . The CONTROL study (ClOsure device iN TRansfemoral aOrtic valve implantation) multi-centre study included 3138 consecutive percutaneous transfemoral TAVI patients, categorized according to vascular closure strategy: Prostar XL vs Perclose ProGlide-based vascular closure strategy . Propensity score matching identified 944 well-matched patients. Composite primary endpoint of major vascular complications or in-hospital mortality occurred more frequently in the Prostar group when compared with the ProGlide group (9.5% vs 5.1%, p=0.016) and was driven by higher rates of major vascular complication (7.4% vs 1.9%, p<0.001) in the Prostar group. In hospital mortality was similar between the groups. Prostar use was associated with higher rates of major bleeding (16.7% vs 3.2%, p<0.001), acute kidney injury (17.6% vs 4.4%, p<0.001) and with longer hospital stay (median 6 vs 5 days, p=0.007).
A recent retrospective study evaluated major and minor bleeding complications as defined by VARC following balloon aortic valvuloplasty in 930 patients . Vascular closure was performed with Angioseal in 69% and with Perclose in 31% of cases. The size of sheath was 9Fr in the majority of cases (84.1%). The Angioseal group had a higher rate of small haematomas (6.9% v 3.5%, p=0.042) but fewer blood transfusions (3.5% v 6.6%, p=0.034). There were no significant differences in in-hospital MACCE and 30 day survival between both devices.
The FemoSeal™ (Terumo Medical Corporation, NJ, USA) is an invasive closure device designed to induce quick and effective haemostasis (FemoSeal\u2122 device.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-10.pptx"})' class="cursor-pointer">Figure 10 and Deployment technique for the FemoSeal\u2122<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-11.pptx"})' class="cursor-pointer">Figure 11). This mechanical device delivers haemostasis without collagen or other sealing agents. FemoSeal™ consists of two resorbable polymer discs, an inner seal deployed intra-arterially and an outer locking disc deployed on the outer wall of the artery. Haemostasis is achieved by sealing the arteriotomy between the two discs, which are held together by a resorbable multifilament complete with friction lock. Haemostasis is achieved once the inner lock is deployed inside the artery. The device automatically signals when you should deploy the outer locking disc, thereby sealing the arteriotomy. The system is resorbed within 90 days. It is indicated when using vascular sheaths sized up to 7 Fr.
  
  FemoSeal™ device.
  
  Deployment technique for the FemoSeal™
The FemoSeal™ vascular closure device was used in the treatment of 1058 patients in the SCAAR registry and was shown to be safe with a bleeding rate of 0.2%.
The CLOSE-UP study was a randomised controlled trial of 1001 patients enrolled in Denmark comparing the Femoseal device with manual compression . The primary endpoint of the study of large groin haematomas (defined as >5 cm) was statistically lower in the Femoseal group (2.2% v 6.7%, p=0.002). Combined endpoint of 14-day adverse vascular events was similar in both groups.
Alonzo et al compared the MACCE and TIMI bleeding rates between transfemoral route sealed with Femoseal and transradial access in a cohort of patients undergoing primary PCI . This was a retrospective and propensity-matched cohort of 777 patients with 511 in the transfemoral and 266 in the transdradial groups. The TIMI bleeding rate in the Femoseal group was 6.6% and statistically higher than that of the radial group (1.3%).
The ExoSeal® (Cordis, Johnson & Johnson, Warren, NJ, USA) VCD is a bioabsorbable device designed for the sealing of femoral artery puncture sites in patients who have undergone diagnostic or interventional procedures using a standard 6 Fr introducer sheath and is indicated for closure of 5F, 6F and 7F arteriotomies. The device achieves haemostasis by means of a visually-guided deployment mechanism which delivers a bioabsorbable polyglycolic acid "plug" atop the femoral artery, anchored by the neurovascular bundle sheath. The plug, which is entirely extravascular, is subsequently hydrolysed into CO2 and H2O via the Krebs cycle over a 3-month period. The device is not suitable for use through sheaths that are longer than 12cm and should not be used in vessel of diameters <5mm.
  
  Deployment technique for ExoSeal® 1. The system is introduced directly into the sheath until the black marker is reached (Figure 12A and Figure 12B).2. The system is clipped into the sheath. This is followed by the appearance of a pulsatile flow of blood through the bleed-back system (Figure 12C).3. Withdraw the system slowly watching the two black and white markers on the upper surface of the system. 4. Initially the flow of blood stops (Figure 12D), then the black and white marker turns entirely black (Figure 12E).5. The system is released by pressing the green button.6. Finish with a non-occlusive dressing for 2-3 minutes.
The pivotal trial of the ExoSeal® vascular closure device was the ECLIPSE trial. This was a 2:1, randomised non-blinded multicentre US trial involving 401 patients comparing the ExoSeal with manual compression. The procedural success rate was 91% with no statistical difference between the groups. Time to haemostasis (4.4 vs. 20.1 minutes) and time to ambulation (2.54 vs. 6.24 hours) were both statistically in favour of the ExoSeal device. There were no major vascular complications .
A 1:1 randomised trial comparing the Proglide with the Exoseal enrolled 100 patients undergoing PCI and endovascular peripheral procedures with end points of immediate total haemostasis and incidence of vascular complication. Immediate haemostasis was more frequent and vascular complication less frequent with the Exoseal device but this did not reach statistical significance .
In contrast, other study found a statistically higher incidence of vascular complications with this device when compared to either Angioseal or Perclose (3.6% v 1.2%) . Logistic regression analysis showed that the use of the Exoseal device was associated with a three fold odds of complications. Repeat angiography performed a month after deployment of the Exoseal device shows angiographic irregularities in nearly 7% of patients treated .
The Mynx® Vascular Closure Device (Cardinal Health, Dublin, OH, USA) is an extravascular vascular closure device (VCD) whose deployment system is designed to minimise the discomfort commonly associated with closing the femoral artery following a catheterisation procedure. The Mynx® received FDA Premarket Approval (PMA) in May 2007.
The Mynx® uses a soft, sponge-like sealant material to close the arteriotomy (Mynx\u00ae vascular device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-13.pptx"})' class="cursor-pointer">Figure 13) The sealant works by rapidly absorbing blood and fluids around the puncture site, swelling in size and covering the hole. This seals the hole and stops the bleeding.
  
  Mynx® vascular device
The sealant material consists of polyethylene glycol (PEG), a water-soluble, non-thrombogenic, conformable, bio-inert polymer which is fully resorbed by the body within 30 days. The Mynx® device uses an extravascular method of deployment. It avoids cinching or tugging of the artery by placing the Mynx® sealant on the outside of the artery. The Mynx® uses the existing 5, 6 or 7 Fr procedural sheath, eliminating the need for a sheath exchange. This avoids potential tissue tract expansion, which can cause tissue trauma and oozing, often resulting in longer bed rest for the patient and potentially delaying ambulation and discharge.
As the Mynx® sealant is placed on the outside of the artery, it is exposed to blood and subcutaneous fluids. The sealant rapidly expands, covering the hole in the artery and conforming to the tissue tract, producing haemostasis. This extravascular placement of the sealant avoids leaving behind an intravascular component, which can compromise blood flow and, in rare circumstances, embolise, requiring surgical repair. The Mynx® sealant avoids initiating platelet activation or an abnormal inflammatory response, and does not generate fibrous scar tissue. In addition, the porous characteristics of the sealant create a platform for natural clot formation and facilitate tissue healing.
The Mynx® VCD includes a balloon catheter with integrated sealant and a 10 ml lock syringe.
  
  Stages in the deployment of the Mynx® device (see text for panel detail)
The current commercially available devices are the MynxAce (MynxAce\u00ae vascular closure device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-15.pptx"})' class="cursor-pointer">Figure 15) and MynxGrip (MynxGrip\u00ae vascular closure device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-16.pptx"})' class="cursor-pointer">Figure 16) vascular closure devices.
  
  MynxAce® vascular closure device
  
  MynxGrip® vascular closure device
The MynxAce device combines the deployment system and extravascular sealant as described above. MynxGrip adds the proprietary Grip Technology to the distal end of the original Mynx Sealant. The result is a sealant that adheres to and seals the arteriotomy while expanding to fill the tissue tract.
This device was studied in a prospective multicentred single arm clinical investigation at 5 European centres. 190 patients were included with 50% undergoing a diagnostic procedure and 50% intervention with sheath sizes 5-7 Fr. The device was found to have a major complication rate of 0.5% in this low to intermediate risk group and no device precipitated complications associated with clinical sequelae were reported .
However other registries have found a significant incidence of pseudoaneurysms , vascular complications , and device failure . These concerns led to a large scale analysis of the NCDR Cath PCI registry to assess the safety of this device . The report showed that the Mynx device was associated with a significantly greater risk of vascular complications (1.2% v 0.8%, p< 0.001), access site re-bleeding (0.4% v 0.3%, p=0.001) and need for transfusion (1.8% v 1.5%, p<0.001) than alternative VCDs. In Manufacturer and User Facility Device Experience (MAUDE) database, 236 reports of Mynx device failure were recorded in the U.S.A from July 2015 to June 2021. The most frequent modes of failure were detachment of the device (46%, n=109), failed deployment (29%, n=68), mal-positioning (13%, n=31), and entrapment (3%, n=8) .
The Femoral Introducer Sheath and Haemostasis (FISH™) Device (Morris Innovative Research, Bloomington, IN, USA) is used to stop bleeding at a puncture site following 5, 6, or 8 Fr cardiac catheterisation procedures. It is a patching device using porcine biomaterial labelled as small intestinal submucosa (SIS) attached to the introducer sheath. Theoretically, the presence of the patch at the arterial entry site during the procedure favours early healing. Following catheterisation, a release wire is pulled detaching the patch from the sheath. The vessel wall is remodelled within 30 days. The FISH™ device also allows for the option of delayed sheath removal in the recovery ward and discontinuation of anticoagulants. The FISH™ device can be used as a combined procedural sheath and closure device, or placed after the procedure and used as a closure device only (The Femoral Introducer Sheath and Haemostasis (FISH\u2122) Device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-17.pptx"})' class="cursor-pointer">Figure 17).
  
  The Femoral Introducer Sheath and Haemostasis (FISH™) Device
The FISH™ closure device was studied in a randomised controlled multicentre clinical trial with almost 300 patients. The study included patients undergoing diagnostic or therapeutic coronary or peripheral procedures performed via the common femoral artery. The procedures were done with an 8 Fr or smaller sheath size. Patients were randomised in a 2:1 fashion to the FISH device or manual compression with primary endpoints of haemostasis and time to ambulation. The results showed a significantly shorter time to haemostasis and mean time to ambulation with the FISH device. There were no significant differences in adverse vascular events. However, it should be noted that 27 patients converted from the FISH device to manual compression .
There have been no studies comparing the FISH device to the market leaders to date.
The StarClose® (Abbott Vascular, Redwood City, CA, USA) is a clip-mediated closure device. It was approved by the Food and Drug Administration in 2005 but has been successfully used in Europe since 2004. In January 2007, StarClose® was approved for interventional use. The StarClose® introduces a small, circumferential, flexible clip, which mechanically binds the surface of the femoral artery together (Position of the nitinol StarClose\u00ae clip on the outside of the artery (there is no intravascular portion<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-18.pptx"})' class="cursor-pointer">Figure 18). The clip is made of nitinol, a nickel titanium alloy whose superelastic properties allow it to return to its original shape once released from the device. Its use involves a 4-step or click process: (1) replacement of the procedural sheath with the StarClose® sheath, (2) deployment of the vessel locator, (3) delivery of the clip, and (4) deployment of the clip, achieving haemostasis. Because the clip is on the outside of the artery, nothing is left behind on the inside to cause potential blockages later.
  
  Position of the nitinol StarClose® clip on the outside of the artery (there is no intravascular portion
  
  Connect the StarClose exchange sheath to the clip applier – “click 1”
  
  Deploy the locator wings – “click 2”
  
  Splitting of the sheath and delivery of the clip to the access site – “click 3”
  
  Deploy clip to close the arteriotomy - “click 4”
The pivotal trial demonstrating the safety and efficacy of the StarClose® Vascular Closure System was the Clip closure In Percutaneous procedures (CLIP) study . This multicentre study enrolled 596 patients in 17 US centres who were undergoing diagnostic and interventional procedures. Patients were randomised in a 2:1 fashion to receive the StarClose VCD or standard compression. 208 patients underwent diagnostic angiography only. In this subgroup, there was no difference in the rates of major and minor vascular complications between the StarClose VCD and manual compression. The device was deployed successfully in 94% of patients and successfully reduced the mean time to haemostasis (15.5 vs. 1.5 minutes, p<0.001) and time to ambulation (268 to 163 minutes, p<0.001) . These findings were confirmed in the interventional subset with no differences in the rates of major complications and a trend towards a lower rate of minor complications in the StarClose group .
A smaller prospective study enrolled 71 patients undergoing diagnostic cardiac or peripheral angiography via the femoral approach in 11 US centres. The 30-day outcomes showed 0% of major complications and 3% of minor complications consistent with the findings of the CLIP study .
A recent randomised controlled trial comparing StarClose and Proglide for femoral haemostasis in patients found a statistically significantly higher rate of vascular complications with StarClose (5.6% v 2.2%) .
The Celt ACD® (Vasorum Ltd, Trinitas House, Dublin, Ireland) achieves closure by applying a biocompatible implant to both the inside and the outside of the puncture site in the femoral artery (which is deep to the skin in the groin area) at the level of the arterial wall. It is available in three sizes to fit 5,6 and 7Fr sheaths. This device does not require exchange of the femoral sheath before deployment (CELT\u00ae vascular closure system<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-20.pptx"})' class="cursor-pointer">Figure 20). Its metallic structure makes it fluoroscopically detectable and thus allows easy detection of mal-deployment. There are some case papers reporting intravascular distal embolization of the system .
  
  CELT® vascular closure system
The CELT ACD trial compared the CELT-ACD against manual compression for femoral site haemostasis in 207 patients undergoing coronary procedures using a 6Fr sheath . The primary endpoint was time to haemostasis while the secondary endpoint was access site complication rate at 30 days. The time to haemostasis was significantly lower in the device group (0 v 8 mins, p< 0.0001) while the 30 day vascular complication rates were similar in both groups.
The Cardiva Catalyst® (Cardiva Medical, Mountain View, CA, USA) is a deployable nitinol disc sheathed in a biocompatible membrane, intended to promote haemostasis at an arteriotomy site as an adjunct to manual compression (Cardiva Catalyst\u00ae II and Catalyst III<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-21A.pptx"})' class="cursor-pointer">Figure 21A). Cardiva Catalyst® is indicated for use in patients undergoing diagnostic and/or interventional femoral artery catheterisation procedures using up to 7 Fr sheaths.
  
  Cardiva Catalyst® II and Catalyst III
The Cardiva Catalyst® consists of a sterile disposable Catalyst® wire and a sterile disposable Catalyst Clip. The wire is available in two versions – Cardiva Catalyst® II and III. The Catalyst II wire contains two agents (kaolin and chitosan) which, when exposed to the tissue tract, accelerate haemostasis by stimulating coagulation, platelet adhesion, and platelet aggregation. The Catalyst III wire has an additional protamine coating which, when exposed to a tissue tract, provides rapid haemostasis in patients anticoagulated with heparin. Catalyst III is indicated for heparinised patients undergoing femoral percutaneous interventions.
The Cardiva Catalyst® still needs assisted manual compression once the disc is collapsed and retracted out of the vessel.
  
  Deployment technique for the Cardiva Catalyst®
This device has been studied in 96 patients following diagnostic cardiac catheterisation, including 25 patients (26%) with contraindications to conventional closure devices. Femoral angiography was performed prior to deployment of the Cardiva closure device. Patients were ambulated at 1 hour after haemostasis was achieved. Successful deployment was achieved in 99% of patients with no major complications and a minor complication rate of 5% .
The Boomerang trial was a prospective, multicentre, randomized, controlled trial comparing the use of the Boomerang wire (now Cardiva Catalyst®) in conjunction with manual compression versus manual compression alone to achieve haemostasis in diagnostic and interventional procedures via the femoral approach . Subjects were randomized 3:1 to Boomerang versus manual compression. No minor or major device-related adverse events were reported. Non-device related complications were 3 (0.9%) in the Boomerang arm (n=327) and 1 (0.8%) in the manual compression arm (n=123). Mean time to haemostasis and time to ambulation were significantly reduced in the Boomerang arm for both diagnostic and interventional procedures (p<0.0001).
The Vascade® Vascular closure system is the third generation of extravascular closure device from Cardiva Medical (VASCADE\u00ae vascular closure system<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-22.pptx"})' class="cursor-pointer">Figure 22). Indicated for femoral arterial access site closure in patients who have undergone diagnostic or interventional endovascular procedures using a 5F, 6F or 7F procedural sheath, it reduces the time to haemostasis and ambulation.
  
  VASCADE® vascular closure system
Vascade® combines both mechanical and biochemical methods of haemostasis for closure of the femoral arterial access site. The main difference with the third generation device is the addition of a protamine sulphate component on the coating applied to the extravascular anchor element in addition to the two other proprietary coatings in the Catalyst® II. The main advantages of the device are a lack of biologic material or permanent implant, as well as the ability to apply it to a wide variety of anatomical conditions .
The safety and effectiveness of Vascade® Vascular Closure System (VCS) was evaluated in 420 patients across 21 clinical sites up to 30 days post diagnostic or interventional procedures in the RESPECT trial with patients randomised 2:1 to VASCADE versus manual compression . There was a high proportion of bivalirudin use in both groups being slightly higher in the Vascade group. (77% vs 69%). The clinical data show that patients treated with the Vascade VCS had a lower mean time to haemostasis and ambulation than the corresponding times for those patients treated with manual compression, and that the differences in these times are statistically and clinically significant. In addition, diagnostic patients treated with the Vascade VCS had a lower mean time to discharge eligibility than the corresponding times for those patients treated with manual compression, and the difference in these times is statistically and clinically significant. The primary safety endpoint was the combined rate of major complications within 30 ± 7 days following the catheterization procedure. The secondary safety endpoint was the combined rate of minor complications within 30 ± 7 days following the catheterization procedure. There were no major complications in the pivotal study in either arm. Additionally, for the total patients the rate of overall minor complications was statistically significantly less in the Vascade treatment arm as compared to manual compression. From October 2013 through March 2020, post-marketing surveillance data (FDA MAUDE database) reported no less than 201 cases of major complications due to VASCADE devices. .
The Axera® device (Arstasis, Redwood City, CA) is the second-generation device from this company and is a single use system for the provision for vascular access and also to promote haemostasis of the vascular site as an adjunct to manual compression (The Axera\u00ae closure device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-23A.pptx"})' class="cursor-pointer">Figure 23A). The system employs a strategy of preclosure and utilises a novel strategy whereby a controlled preprocedure arteriotomy results in an overlap of arterial tissue after sheath removal that is reinforced by hydrostatic arterial pressure that creates closure. This approach leaves no foreign material behind. The Axera device is approved for used in diagnostic cases with 5F or 6F sheaths.
  
  The Axera® closure device
The Arstasis™ family of devices uses a traditional femoral entry approach. Following the needle stick and wire insertion, the Axera® is deployed in 3 steps (Access device introduction<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-23B.pptx"})' class="cursor-pointer">Figure 23B, Self-sealing arteriotomy<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-23C.pptx"})' class="cursor-pointer">Figure 23C and Device removal<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-23D.pptx"})' class="cursor-pointer">Figure 23D).
  
  Access device introduction
  
  Self-sealing arteriotomy
  
  Device removal
The 0.035” Latchwire is introduced through the access needle and the needle then removed. The Latchwire is attached to the distal top of the Axera® device. The device is then advanced into the artery until blood is observed through the marker port.
The heel actuator is slid back to deploy the heel inside the artery. The Axera® is then retracted at 55-60° until resistance is felt and the blood mark stops or diminishes. The plunger is depressed to advance the micropuncture needle and create a self-sealing arteriotomy. The 2nd mark should be observed through the marker port.
The guidewire is introduced through the plunger port and the micropuncture needle. The plunger is retracted to withdraw the needle and the heel released. The device is removed. The sheath, dilator and dilator adaptor assembly is then advanced. The sheath may be removed when clinically indicated. Brief manual compression, above the puncture site, is then applied until haemostasis is achieved.
The RECITAL study was a prospective multi centred study which enrolled 351 patients to assess device and procedural success of key post-catheterization recovery measures that influence the safety and speed toward discharging those patients from the hospital. Key results of the RECITAL study showed that device deployment was successful in 97% of RECITAL patients and femoral bleeding stopped after only 3 minutes for diagnostic angiogram patients and 5 minutes for anti-coagulated patients undergoing a percutaneous coronary intervention (PCI). Most of the patients were able to sit up within 15-30 minutes and were able to walk in about one hour after an angiogram and two hours after a PCI. Investigators found no major complications related to use of the Arstasis device.
A small histological study by Patel et al involved examining bovine arteries following AXERA® and control (Seldinger access) arteriotomy and indicated that healing after AXERA® access was similar to Seldinger access technique with no evidence of dissection or pseudoaneurysm .
Extravascular manual compression, with or without the potentially augmenting effect of haemostasis patches, remains the gold standard of vascular sheath removal. The technique involves applying digital pressure proximal to the skin puncture point in order to compress the femoral artery and surrounding soft tissues against the femoral head. Vertical pressure through a straight upper limb reduces the operator fatigue that can be encountered. Compression should be applied such that haemostasis is achieved with a palpable pulse. If this is difficult to achieve (e.g., because of large body habitus), then sufficient compression can be applied to obliterate temporarily the femoral pulse (10 minutes on/off). Some operators reduce the blood pressure with nitrate medication in significantly hypertensive individuals in order to facilitate haemostasis.
An arterial puncture, which is either proximal or distal to the femoral head, means manual compression will generate less pressure over the puncture point and hence be less successful in achieving optimal haemostasis.
Although manual pressure is widely used, especially for smaller diagnostic sheaths, the technique is inexact and often stressful to the patient and operator, should haemostasis be difficult to achieve. Interventional procedures usually require the use of larger sheath sizes and are performed in combination with an anticoagulation regime. For these patients, once the ACT has dropped to a satisfactory level, manual compression can be used to control bleeding. Therefore, early sheath removal is precluded and bed rest of a variable duration is necessary, dependent on the level of anticoagulation. These circumstances may exist with excessive bleeding, intra-aortic balloon pump (IABP) removal, or arterial insufficiency.
The use of a mechanical compression device in these and many other scenarios can be beneficial in achieving primary haemostasis with minimum discomfort to the patient and at a lower cost than a closure device.
In addition, for certain high risk patients who have received aggressive anticoagulation therapy, indwelling sheaths have to be left in situ for several hours or overnight and compression must be maintained for an extended period of time after sheath removal. Prolonged pressure bandage is also a source of discomfort and hazard since the development of complications such as haematoma, oozing and arterial bleeding are concealed and noticed much too late.
It should be noted that there are several situations where femoral anatomy is non-ideal (a small calibre vessel < 5 mm, a calcified artery by fluoroscopy, PVD up to 70% stenosis, and arteriotomy site at or below the bifurcation in diagnostic catheterisation procedures) and which preclude use of a vascular closure device. In these circumstances, manual compression should remain the closure method of choice.
The FemoStop™ Gold Femoral Compression System (Abbott Vascular, Santa Clara, CA, USA) is indicated for use in the compression of the femoral artery or vein after vessel cannulation, and in ultrasound guided compression repair after femoral artery pseudoaneurysm. This device is, however, contraindicated in patients with severe peripheral vascular disease.
The system consists of an arch with a sterile pneumatic pressure dome, a belt and reusable pump with manometer (FemoStop\u2122 compression device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-24.pptx"})' class="cursor-pointer">Figure 24). The pressure dome is placed over the vessel puncture site in the groin. The belt is placed around the patient. The dome, when inflated, applies a mechanical pressure over the vessel puncture site to induce haemostasis. The reusable pump and the manometer control the pressure of the dome and allow the application of the desired pressure to maximise haemostasis. The arch and the belt provide counter-pressure for the dome. The transparency of the dome allows for total puncture visibility. Importantly, the point of maximal pressure identified at the apex of the dome must be applied proximal to the femoral artery puncture. A potential error is to assume that the skin incision/puncture point represents the femoral arterial puncture point.
  
  FemoStop™ compression device
Clinical studies of the use of passive devices are limited especially as the FemoStop™ has been used primarily as the control arm in studies after endovascular procedures . A small randomised study found no differences in groin complications with FemoStop when compared with manual compression . By contrast, a randomised study which enrolled 212 patients showed FemoStop use to be associated with an increased rate of vascular complications when compared with manual compression .
The Safeguard® (Maquet Cardiovascular [Datascope Corp.], Fairfield, NJ, USA) device is a latex-free sterile pressure-assisted dressing which is designed to maintain and protect haemostasis while providing the ability to monitor clearly the access site. Safeguard® is a single-use adhesive dressing with a transparent inflatable plastic bulb (Safeguard\u2122<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-25.pptx"})' class="cursor-pointer">Figure 25). When inflated, the bulb is inelastic and provides consistent pressure over the wound site to maintain haemostasis. The see-through bulb allows easy monitoring of the wound site. The Safeguard® is available in 12cm and 24cm sizes.
  
  Safeguard™
Haemostasis pads contain pro-coagulant coatings that serve to enhance the coagulation pathway with the aim of achieving quicker vascular repair. Several devices are available on the market including the Clo-Sur P.A.D® (Merit Medical Systems, Inc., South Jordan, UT, US), D-stat® Dry (Vascular solutions, MN, US), Neptune Pad (TZ Medical Inc., Portland, OR, US) and SyvekExcel® Patches (Marine polymer technologies Inc, Dankers, MS, US). Their mode of action is illustrated by the Clo-Sur P.A.D® below.
The Clo-Sur P.A.D® (pressure applied dressing) (Merit Medical Systems, Inc., South Jordan, UT, US) device is a soft, non-woven hydrophilic wound dressing. This device consists of the hydrophilic naturally occurring biopolymer Polyprolate™ acetate. This linear biopolymer is cationically charged in its dry state. It is the chain of positive charges which endows Polyprolate™ with its blood-coagulant properties. Cationic Polyprolate™ forms a coagulum with heparinised blood, defibrinated blood and washed red blood cells.
Studies of haemostasis pads as an adjunctive technique to manual compression compared to compression alone have generally demonstrated limited efficacy with similar time to haemostasis, vascular complications rate and bed-rest time , . There is also a high technical failure rate associated with these devices .
The largest analysis of these devices comes from the National Cardiovascular Data Registry, which showed that their use is associated with a slightly smaller risk of bleeding, or vascular complications compared to conventional manual compression .
The topic of vascular closure following TAVIs is covered in chapter Transcatheter aortic valve implantation (old) but it is worth briefly mentioning some of the devices available for the closure of large arteriotomies. Structural heart procedures such as TAVI and the implantation of mechanical circulatory support devices (e.g Impella) require large bore sheaths (between 14-20Fr). Larger arteriotomies are associated with greater morbidity including higher bleeding risk, delayed ambulation as well as higher cost of care. Surgical closure offers the best control of bleeding but is associated with a higher incidence of infection and need for debridement .
Closure of large bore access sites can be challenging using conventional VCDs. The currently widely utilised technique is that of the preclosure with either the Prostar XL and the Proglide VCD. This is done by partially deploying the sutures after the artery is accessed by a small calibre sheath and then replacing the smaller sheath with the large-bore arterial sheath (Preclosure technique using two Proglides prior to TAVI insertion.<\/p>","mediaType":"video","mediaMimeType":"video\/mp4","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_VIDEO-1.pptx"})' class="cursor-pointer">Video 1). After completion of the procedure, the sheath is removed and the sutures tied in the standard manner. However this technique is associated with a not inconsequential bleeding rate . A retrospective analysis of 472 propensity matched patients undergoing transfemoral TAVI showed that the Prostar VCD was associated with a higher major vascular complication rate .
Preclosure technique using two Proglides prior to TAVI insertion.
The exponential growth of TAVI over the past decade has spurred interest in the development of dedicated VCDs for large bore arteriotomy closure. Three of the devices available on the market are briefly outlined below.
The MANTA™ device (Essential Medical Inc, Malvern, Pennsylvania) has CE mark to close puncture sizes ranging from 10 to 25 French sheaths. The concept is similar to that of the Angio-seal. The device uses a collagen implant which is secured by an intra-luminal anchor that allows closure of large-bore arteriotomies (MANTA\u2122 vascular closure system<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-26.pptx"})' class="cursor-pointer">Figure 26). The device is available in two sizes – 14 and 18Fr which close 10-14 Fr and 15-22 sites respectively.
  
  MANTA™ vascular closure system
The PerQseal® device (Vivasure Medical Ltd, Galway, IR) is a fully bioresorbable device indicated for closure of TAVI access sites. The device consists of a polymer implant which is deployed inside the vessel wall (PerQseal\u00ae vascular closure system<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-27.pptx"})' class="cursor-pointer">Figure 27). The implant has a flexible intra-vascular patch and a supporting scaffold. Its abluminal surface is textured to promote adherence of the patch to the vessel wall. Part of the scaffold extends through the arteriotomy and includes a locator which helps maintain the implant in position. After deployment, the implant is endothelialised and fully absorbed.
  
  PerQseal® vascular closure system
The Inseal VCD (Inseal Medical Ltd, Caesarea, Israel) is indicated for closure of punctures ranging from 14-21Fr sizes. The device consists of a self-expandable Nitinol frame covered by a biodegradable membrane. This membrane covers the puncture site and is maintained in place by the blood pressure (Inseal vascular closure device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-28.pptx"})' class="cursor-pointer">Figure 28). This system is delivered through the same sheath used for the large bore procedure.
  
  Inseal vascular closure device
Diagnostic and interventional cardiovascular procedures are routinely performed via the common femoral or the radial artery. The transfemoral route initially dominated the interventional landscape but has largely been superseded by the radial approach in many countries in Europe and Asia. While the transfemoral route is still widely used in United States, data from the National Cardiovascular Data Registry shows a growing adoption of the radial approach . Transradial access is currently the standard vascular access site for percutaneous coronary procedure.
Transradial access is associated with fewer bleeding and vascular complications , , , . This can be mostly explained by the fact that the radial artery is a superficial and easily compressible vascular site. Consequently, most of the disadvantages of the femoral technique are not encountered with the transradial approach. Even in obese patients, the radial artery is close to the skin surface, making the initial needle puncture simple and straightforward. For the same reason, when the procedure has been completed, a short compression of the radial artery can stop the bleeding, even when the patient has been aggressively anticoagulated. Should any bleeding occur, it can be seen immediately. Finally, unlike the proximity of the femoral artery to the femoral nerve, the radial artery is not close to a major nerve, so the likelihood of nerve injury during the procedure is very low.
A meta-analysis of four randomized and seven non-randomized studies, which compared the radial approach with the femoral approach plus achievement of haemostasis with VCD, showed a reduction in both access site complications and major bleeding when the radial approach was used . Furthermore, this benefit has also been observed/ in the context of acute coronary syndromes. In a large-scale meta-analysis of 18 randomized controlled trials, trans-radial procedures were associated with decreased all-cause mortality in patients with ACS . Another systematic review of six prospective and eight retrospective studies (radial access, n = 8032; femoral access, n = 23 031) has demonstrated the relevance of the radial access in the setting of STEMI complicated by cardiogenic shock .
In addition to lower complication rates, another advantage of the radial technique is that patients do not need to lie flat and still for 4-6 hours, or experience what is sometimes a painful manual compression of the artery. Patients leave the catheterisation lab and are able to sit up and walk almost immediately. Because of the simpler healing process for the arterial puncture in the wrist, patients may also be discharged home without having to spend the night in hospital, which impacts the cost-effectiveness of interventional procedures.
The 2018 ESC/EACTS guidelines on myocardial revascularisation followed by the 2020 guidelines on the management of ACS-NSTEMI, recommend the radial approach as the standard access of choice for PCI unless not technically feasible , . Similarly, a recent position statement from the AHA recommends the transradial approach as the default strategy in patients presenting with ACS .
As the use of the transradial access has increased, so has the demand for compression devices to address this growing need. Manual pressure with gauze were initially replaced by rigid wrist bands such as Hemoband. Nowadays, these have largely been superseded by air filled bracelets.
A number of devices have been designed to provide haemostasis, from a simple, plastic haemodialysis band compressing a gauze bullet over the radial arterial puncture to specifically designed compression bands (Radial closure devices (non-exhaustive)<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_TABLE-4.pptx"})' class="cursor-pointer">Table 4). Topical haemostatic agents are also available for use in conjunction with radial bands in order to shorten the time to achieve haemostasis. A wide variety of compounds are available for this purpose. It is highly recommended that just enough pressure should be applied to prevent bleeding but not so much as to cause complete collapse of the vessel under pressure resulting in flow cessation.
  
  Radial closure devices (non-exhaustive)
Regardless of the chosen device, there are a few critical steps in this process. After removing the drapes from the hand and arm, begin sheath removal as noted below.
The patient is given instructions about puncture site compression with the fingers if (rare) late bleeding occurs.
Two of the commonly used compression devices are outlined below.
This is a radial artery compression device designed to stop bleeding safely and comfortably after transradial procedures (TR band\u00ae. The green locator dot must be positioned proximal to the puncture site<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-29A.pptx"})' class="cursor-pointer">Figure 29A and TR band applied to the wrist<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-29B.pptx"})' class="cursor-pointer">Figure 29B) The TR Band® (Terumo Medical Corp., Somerset, NJ, USA) is an easy-to-apply transparent band with precision pressure balloons which enable unobstructed visualisation and monitoring of radial artery compression.
  
  TR band®. The green locator dot must be positioned proximal to the puncture site
  
  TR band applied to the wrist
The primary benefit that the TR Band® has over similar devices is its directed radial compression, which enables haemostasis achievement without compromising local nerve structure. The TR Band® does so with its two separate inflatable balloons. A large balloon compresses the entire puncture site, and a small balloon gives it an angled direction for point compression. This dual system enables the TR Band® to avoid compression of the nerves or the ulnar artery, which can cause discomfort and/or complications. The transparency of the band, which allows full visibility to the clinician during the critical post-procedure stage, is another significant clinical benefit of this product.
The Helix™ device (Vascular perspectives) allows a controlled compression of the radial artery (Helix\u2122 radial compression device<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_FIGURE-30.pptx"})' class="cursor-pointer">Figure 30). The aim is to improve radial patency rate post procedure. The design features a screw-bolt on top of the transparent silicone pressure pad. The steps are outlined below:
  
  Helix™ radial compression device
Over the last few years, a new technique of accessing the distal radial artery via the anatomical snuffbox has been applied . It was first reported in 2011 for the retrograde opening of occluded radial arteries and then again in 2017 as a more comfortable technique for left radial access. In addition, post-procedural radial artery occlusion being a common complication of the fore-arm radial access, this technique helps mitigate the risk of radial artery occlusion proximal to the wrist. It is, therefore, gaining popularity after demonstrating greater comfort for both patients and operators, effectiveness and decreased occurrence of vascular occlusion . For instance, patients with specific orthopaedic limitations, who are unable to supinate their arm, can benefit from this technique. More importantly, it preserves the use of the radial artery for an eventual coronary bypass grafting, or an arteriovenous fistula. And yet, in the DISCO RADIAL trial, the first large-scale multicentre controlled randomized trial comparing DRA to TRA, there was no difference in the radial occlusion rates when a rigorous haemostasis protocol was applied .
As the traditional radial access, due to its location directly above the carpal bones, it has the advantage of being easily compressed to achieve an effective haemostasis. A dedicated device has recently been developed “The PreludeSYNC DISTAL” radial compression device ((PSD; Merit Medical Systems, Inc., South Jordan, UT).
Specifically designed for distal radial compression, this device has a soft wristband and thumb saddle strap with secure hook and loop fasteners. The band delivers adjustable compression of the puncture site with an inflatable balloon, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement. The main steps of its use are outlined below:
VCD pivotal studies have generally included highly selected, low-risk patients in studies with small sample sizes, with 250 to 600 patients. Complication rates of 3-5% have been reported with use of VCDs , , , . Without the inclusion of higher risk patients, there is a multitude of precautions and contraindications for each device (Contraindications and precautions for VCD use*Instructions for Use: Angio-Seal VIP, St. Jude Medical, February 2006 (www.sjm.com); \u03c8 Instructions for Use: Perclose AT (August 2006) and StarClose (February 2007), Abbott Vascular (www.abbot-tvascular.com\/ifu).GP: glycoprotein; HTN: hypertension; NM: not mentioned in instructions for use as warning or caution; PCI: percutaneous coronary intervention; PVD: peripheral vascular disease; SFA: superficial femoral artery; VCD: vascular closure device; +: explicitly mentioned in instructions for use as warning or caution.<\/p>","mediaType":"image","mediaMimeType":"image\/jpeg","mediaPPT":"PCR_Textbook_Chapter-VASCULAR-CLOSURE_TABLE-5.pptx"})' class="cursor-pointer">Table 5)
  
  Contraindications and precautions for VCD use*Instructions for Use: Angio-Seal VIP, St. Jude Medical, February 2006 (www.sjm.com); ψ Instructions for Use: Perclose AT (August 2006) and StarClose (February 2007), Abbott Vascular (www.abbot-tvascular.com/ifu).GP: glycoprotein; HTN: hypertension; NM: not mentioned in instructions for use as warning or caution; PCI: percutaneous coronary intervention; PVD: peripheral vascular disease; SFA: superficial femoral artery; VCD: vascular closure device; +: explicitly mentioned in instructions for use as warning or caution.
There have been many case reports of rare complications occurring with VCDs such as infections, arterial laceration, uncontrolled bleeding, pseudoaneurysm, arteriovenous fistula, device embolism and limb ischaemia , . Whilst complications are rare, they can be devastating, resulting in either surgical repair, often under urgent circumstances, or even limb loss. These data underline the importance of meticulous attention both to the suitability of a femoral puncture for device closure and to the technique itself, which requires formal supervised training. Any persisting discomfort following device deployment should prompt a thorough examination and imaging to exclude complications.
As a result of many adverse events post-VCD deployment, the FDA initiated a study of 166,680 patients via the ACC-NCDR database “to assess the relative risk of complications after the use of the 2 main types of haemostasis device” . However, concerns that use of VCDs led to worse outcomes than manual compression have not been supported by multicentre registries and meta-analyses.
Given the availability of many randomised studies with conflicting safety data and multiple limitations, it has become important to consider the evidence from larger registries and databases regarding the efficacy and safety of the devices as well as their comparisons against manual compression and against each other. Some of the larger trials and registry data are outlined below.
A large single-centred retrospective analysis compared vascular outcomes in almost 6,000 patients treated with manual compression with 516 VCD uses (371 Angio-Seal, 101 Techstar [suture-based]) after diagnostic and interventional procedures. This demonstrated that the use of VCDs was associated with significantly more haematomas, a larger drop in haematocrit and a greater need for vascular surgical repair of the access site. Multivariate analysis was performed and clarified that age, body surface area and sex, not the use of VCDs, were independent predictors of vascular complications. These data reflect the early experience with VCD and their interpretation should take into account the “learning curve” of the operators.
The Acuity trial compared access site bleeding in 11,621 patients undergoing diagnostic and interventional procedures. 4,307 of these patients (37.1%) received a VCD . Rates of access site bleeding were lower with VCD compared with no VCD (2.5% vs. 3.3%, p=0.01). Stepwise logistic regression revealed that a VCD was an independent determinant of freedom from major access site bleeding. The ISAR-CLOSURE trial compared a VCD-based strategy (with Femoseal or Exoseal) against manual compression to close arteriotomies after diagnositic coronary angiography in 3015 patients . The primary endpoint of access site-related vascular complications at 30 days was similar in both groups (6.9% with VCD and 7.9% with compression).
More recently, in a post-hoc analysis of ISAR-CLOSURE trial , 3018 patients who underwent transfemoral diagnostic coronary angiography, were randomly assigned to arteriotomy closure with FemoSeal or manual compression. Vascular access-site complications were lower in the FemoSeal VCD group compared to manual compression (6.0% vs 7.9%; P=.04). Time to hemostasis was significantly shortened with the FemoSeal VCD compared to manual compression (0.5 min [IQR, 0.2-1.0 min] vs 10 min [IQR, 10-15 min]; P<.001 .
Angio-Seal and Perclose were compared with manual compression in 1,500 patients with a finding that Angio-Seal was associated with a shorter time to haemostasis and ambulation than Perclose . Perclose use was associated with more access-site complications than Angio-Seal and manual compression. The Angio-Seal and Perclose devices were again compared with manual compression in a registry of patients undergoing PCI with abciximab in which 524 patients used the Angio-Seal, 2177 the Perclose, and 1824 underwent manual compression. Haemostasis was significantly improved in the Angio-Seal group versus the Perclose group (p<0.05) and overall complications were reduced among patients with successful VCD use compared with manual compression (p<0.05). A single centre retrospective study compared five different closure devices (Angioseal, Perclose, Starclose, Exoseal and Mynx) used in an outpatient endovascular surgery centre . Femoral arterial closure with closure devices was performed in 1810 patients. The Angioseal device had the smallest rate of failure (3.5%) and minor complications (1.3%).
In a large-scale, randomized, multicenter trial, a sex-specific analysis was conducted on 1395 women, randomly assigned to arteriotomy closure with an intravascular VCD, extravascular VCD, or manual compression (MC) after transfemoral diagnostic coronary angiography. There was no significant difference concerning vascular access-site complications between the VCD and manual compression groups. Time to hemostasis was significantly shortened with VCD compared with MC (1 [interquartile range, 0.5–2.0] minutes) versus 11 [interquartile range, 10–15] minutes; P<0.001). .
The British Cardiovascular Interventional Society (BCIS) examined the 30 day mortality between manual pressure and VCD in the UK registry of 271845 patients who had undergone PCI between 2006 and 2011 . After propensity matching, the use of VCDs was found to be associated with a small but statistically significant benefit in 30 day mortality (1.8% v 2.0%, p<0.001). This effect was more pronounced in females and acute presentations (ACS).
Data from the large ACC-NCDR registry has been used to compare VCDs versus manual compression. In an analysis of 166,680 patients across 214 sites, the risk of vascular complications was lower in men than in women, and after diagnostic rather than interventional procedures when a VCD was used. Multivariate analysis demonstrated VCD use to be associated with a lower risk of any vascular complication, driven predominantly by lower rates of bleeding and pseudoaneurysm formation in diagnostic procedures only. In a more recent analysis of the same registry, a total of 2056585 cases were examined with just over half of the cases (51.2%) having received a VCD . Global vascular access site complication rate was 1.5%. The use of VCD was associated with a 0.36% absolute risk reduction in vascular access complications compared with manual compression.
In a more recent large multicentre American registry, comprising more than 11000 patients, treated by PCI via the femoral artery between 2015 and 2021, VCD use compared to manual femoral compression was associated with significantly reduced vascular access complications and bleeding events. Among closure devices, Perclose and Angio-Seal showed a lower rate of hematoma than Mynx .
  
  Summary of meta-analyses
Several meta-analysis have been conducted comparing VCDs with manual compression and are briefly discussed below.
Koreny et al looked at data assessing the efficacy and safety of VCDs compared with manual compression in 30 randomised trials enrolling 4,000 patients. Overall time to haemostasis was reduced by a mean of 17 minutes with VCDs. There were non-significant trends towards increased haematomas, bleeding, arteriovenous fistulas and pseudoaneurysms with VCDs .
Nikolsky et al analysed the safety of VCDs versus manual compression in a meta-analysis of 37,066 patients in 30 centres . In patients undergoing PCI, there was an increased rate of vascular complications (OR 1.34, 95% CI 1.01 to 1.79) when compared with manual compression. However, in patients having only a diagnostic procedure, there was a non-significant trend which favoured VCDs over manual compression. When the devices were compared, the Angio-Seal and Perclose devices had comparable results to manual compression; however, the VasoSeal was associated with increased vascular complications compared with manual compression (OR 2.27, 95% CI 1.35 to 3.80).
Vaitkus et al pooled data from 16 randomised controlled trials with 5,084 patients and showed an overall reduction in vascular complications with VCD use (OR 0.89, 95% CI 0.86 to 0.91) , in particular with the Angio-Seal and Perclose devices. The VasoSeal device was associated with a higher risk of vascular complications. When they looked at PCI procedures only, there was a significant reduction in vascular complications with the Angio-Seal (OR 0.51, 95% CI 0.45 to 0.58), no difference with Perclose (OR 1.0, 95% CI 0.13 to 7.48) and an increased risk with VasoSeal (OR 1.18, 95% CI 1.16 to 1.20).
Another meta-analysis evaluated the safety and efficacy of VCDs in 7,528 patients in 31 prospective randomised studies . It is worth noting that most of these studies excluded patients at high risk of puncture site complications. There was no overall difference in terms of groin haematoma, bleeding, pseudoaneurysm and blood transfusion. The use of VCDs was associated with an increased risk of infection (p=0.02), lower limb ischaemia/arterial stenosis/device entrapment in the artery (p=0.07) and need of vascular surgery for arterial complications (p=0.10). However, the use of VCDs was associated with a significantly shorter time to haemostasis and maybe a shorter recovery time.
Cox et al. conducted a systematic review of 34 randomised controlled trials (RCT) comparing vascular closure devices to manual compression . The use of VCDs shortened time to haemostasis, allowed more rapid ambulation and earlier discharge compared to manual compression.
A recently published meta-analysis by Iannaccone et al showed that VCDs did not result in fewer vascular complications compared to manual compression . However they did have a shorter time to haemostasis. In terms of specific devices, the StarClose had the lowest vascular complication rate while the Angioseal device had the shortest time to haemostasis.
A Cochrane review of 52 randomised controlled trials conducted by Roberson et al. showed that both metal clip-based and suture-based VCDs achieved haemostasis more quickly compared to extrinsic compression . Mortality and vascular injury rates between VCDs and extrinsic compression were comparable. The safety and efficacy profiles of the different devices were comparable.
It is difficult to make accurate evaluations as to the efficacy and safety of VCDs from the observational registry data and meta-analyses. The randomised trials are small and appear to have been conducted in patients with lower risk than the current clinical population. These small trials may not accurately identify the different clinical effects of the devices. In addition, the study endpoints are variable with some studies having a composite endpoint , , some using a single endpoint , , and some using a different definition of “vascular complications”.
Taking all this into account, there are several conclusions which may be drawn:
Guidelines published by the ACC/AHA give the following recommendations:
The VCDs with the longest commercial availability are the Angio-Seal™ and the Perclose®. Both have undergone significant modifications and improvements though they are still associated with device failures. Multiple clinical, procedural and treatment factors influence arterial access. Patient size, age, use of glycoprotein IIb/IIIa inhibitors, multiple attempts to attain vascular access and suboptimal punctures will continue to provide a challenge for vascular closure. With the advent of increased radial PCI worldwide, this may further impact on the indications and necessity for VCDs.
Even though the radial access is increasingly being used, femoral access is still necessary in many situations e.g where larger arteriotomies are required for the interventional procedure or in cases where the radial artery is not accessible. Vascular complications remain of major concern with percutaneous femoral access, but may be minimised with a strategic approach to vessel closure. With increasing technological advances over the last decade, the currently available VCDs are associated with increased success rates and a low, yet real risk of vascular events that can be decreased with particular attention. They provide increased comfort for patients and allow earlier ambulation.
For diagnostic and interventional cases performed by femoral approach, it may be reasonable to consider the use of VCDs in most patients in order to decrease complication rates and improve time to ambulation , , . The increasing use of the radial approach may further limit vascular complications, improve time to ambulation and prove to be cost-effective.
The use of femoral closure devices after completion of angiography or PCI has two purposes: the improvement of patient comfort and the reduction of complication rates.
Although the achievement of these objectives depends on the systems used as well as on the level of operator experience, they have become all the more crucial as the transfemoral approach is increasingly challenged by the transradial route which is safer and has even been associated with a decrease in mortality among specific groups of patients.
Many patients treated years ago have bad memories of the completion of the procedure and are frightened of undergoing another intervention; long and painful manual compression, sand bags, and tight compressive bandages have become a thing of the past.
The advent of TAVI and the use of circulatory support systems have made it imperative for interventional cardiologists to master the transfemoral approach, especially as coronary procedures are mostly performed via the radial approach. Indeed, the devices used for TAVI and circulatory support are large, often larger than 12 to 14 Fr, which makes all potential complications even more catastrophic.
The quality of the femoral puncture is essential. Indeed, performing the puncture too high may result in the occurrence of retroperitoneal haemorrhage; if performed too low, it may lead to the formation of a false aneurysm. Transfixing punctures should not be performed at any cost: Closing the anterior side of the femoral artery is of no use when the posterior side is bleeding…
For preference, we recommend that only one or 2 vascular closure systems should be used. Indeed, using the same system frequently enables the operator to master its use, to achieve a high success rate and to anticipate potential complications. Personally, I prefer collagen-free closure systems because, when their use is unsuccessful, the results are immediately observed and the operator is able to use manual compression and reverse heparin when required. After TAVI, a number of experienced operators combine successfully a suture and a collagen device.
In view of my experience with the Prostar and the Perclose devices and having performed thousands of vascular closures, I can confirm that these systems are associated with a high success rate and that, as with the radial approach, patients can be discharged three hours after the procedure with an extremely low rate of local complications.